CRT
|
- Configuration and useful screen : 6 inch
rectangular screen with internal
graticule, 8X10div.(1div = 1cm) marking
for measurement of rise time, 2mm
subdivisions along the central axis. -
Acceleration potential : +1.9kv approx.
(ref. Cathode)
- Phosphor : P31
- Focusing : provided
- Trace rotation : provided
- Intensity control : provided
|
|
Z-AXIS INPUT
(INTENSITY
MODULATION)
|
- Input signal : positive going signal
decreases +5Vp-p or more signal
cases intensity noticeable modulation at
normal intensity setting
- Bandwidth : DC to 2MHz(-3dB)
- Coupling : DC
- Input impedance : 25kΩ - 30kΩ typical
- Maximum input voltage : 30V(DC + peak AC)
|
|
VERTICAL
DEFLECTION
|
- DC coupled(-3dB) : DC to 30MHz normal / DC
to 10MHz magnified
(CH 1 only)
- AC coupled(-3dB):10Hz to 30MHz normal /
10Hz to 10MHz magnified
(CH 1 only)
- Modes : CH1, CH2, ADD, DUAL(CHOP :
Time/div. switch from 0.2s to 1ms,
ALT : Time/div. switch from 0.5ms to
0.2μs)
- Deflection factor (normal): 5mV/div. to
20V/div. in 12 calibrated steps of 1-2-5
sequence. Continuously variable between
steps at least 1:2.5 X 5.
- Deflection factor (5MAG): 1mV/div. to
4V/div. in 12 calibrated steps
(CH1 only).
- Accuracy : normal ±3%, magnified : ±5%(CH
1 only)
- Input impedance : approx. 1MΩ in parallel
with 30pF
- Maximum input voltage : direct 400V (DC +
peak AC)
- Input coupling : AC, GND, DC
- Rise time : 17.5ns or less (35ns or less
in 5MAG)
- CH1 output : 25mV/div. into 50Ω / DC to
10MHz(-3dB)
- Polarity inversion : CH2 only
|
|
HORIZONTAL
DEFLECTION
|
- Sweep mode : NORM, X-Y, X 10, Variable
- Time base A : 0.2μs/div. to 0.2s/div. in
19 calibrated steps of 1-2-5 sequence
uncalibrated continuous control between
steps at least 1:2.5
- Sweep magnification : 10 times(maximum
sweep rate : 20ns/div.) 50ns/div.
or 20ns/div. ±10%
- Accuracy : ±3%(0℃ to 35℃), ±5%(0℃~10℃,
35℃~40℃)
/ additional error for magnifier : ±2%
|
|
TRIGGER
SYSTEM
|
- Modes : AUTO, NORM, TV-V, TV-H
- Source : VERT(DUAL/ALT), CH1, LINE, EXT
- Coupling : AC
- Slope : + or -
- Sensitivity and frequency in AUTO, NORM
20HZ - 2MHz :
0.5div. or 2div.(VERT) in INT/ 0.2Vp-p in
EXT
- Sensitivity and frequency in AUTO, NORM
2Hz - 20MHz :
1.5div. or 3div.(VERT) in INT/ 0.6Vp-p in
EXT
- Sensitivity and frequency in AUTO, NORM
20MHz - 30MHz :
2.5div. or 4div.(VERT) in INT/ 1Vp-p in
EXT
- TV-V, TV-H : at least 1div. or 1.0Vp-p
- External trigger input impedance : 1MΩ ±
10%
- Maximum input voltage : 400V (DC + AC
peak)
|
|
X-Y OPERATION
|
- Sensitivity : same as vertical deflection
for both X-axis(CH1) and Y-axis
(CH2) except accuracy ± 5% in X-axis -
X-axis bandwidt : DC to 500kHz
(-3dB)
- X-Y phase difference : 3° or less as DC to
50kHz
|
|
CALIBRATOR
|
- Probe adjustment : approx. 1kHz
frequency(20%), 0.5V(10%) square wave
with 40-50% duty ratio
|
|
POWER SUPPLY
|
- Line voltage range : 115V(98-125V) with
125V 1A fuse or 230V(198-250V)
with 250V 0.5A fuse - Line frequency :
50/60Hz
- Power consumption : approx. 45W
|
|
PHYSICAL
CHARACTERISTICS
|
- Weight : 7.8kg
- Dimension : 316mm(W) X 143mm(H) X 406mm(L)
|
|
OTHERS
|
- Accessories including : Operation manual,
Spare fuse, Power cord
- Optional accessories (sold separately) :
Probe
|